SY3.5 Mobile High Frequency Veterinary X-ray Machine
Main technical parameters (high frequency)
1 Power requirements:
Single-phase power supply: 220V ± 22V (safety standard socket)
Power frequency: 50Hz±1Hz
(Power capacity): 4kVA
Power supply resistance: <0.5Ω
2 Standard sizes:
Tube horizontal extension: ≥ 800mm
The highest distance from the ground: 1800mm ± 20mm
The lowest distance from the ground: 490mm ± 20mm
Equipment parking size: 1400 × 640 × 1300 (mm)
Equipment quality: 130 (kg)
3 The main technical parameters
Rated output power: 3.2 kW
Tube: Fixed anode tube XD6-1.1, 3.5 / 100
Anodic target angle: 19 °
Limiter: Manual adjustment
Fixed grid: 2.5mm aluminum equivalent (X-ray tube with beam restraint)
Positioning light: 24V halogen bulb; the average illumination of not less than 100 Lx
Maximum cassette size / 1 m SID: 450 mm × 450 mm
The maximum ground slope when moving is ≤10 °
Rated output power: 3.2kW (100kV × 32mA = 3.2kW
Tube voltage (kV): 40 —— 100kV (1kV increase / decrease)
Tube current (mA): 32 mA, 40 mA, 50 mA
Exposure time (s): 0.01 —— 4 s
Current time product (mAs): 1 —— 200mAs
Tube current Tube voltage
32 40——100
40 40——80
50 40——60
4、Features
For hospital wards and emergency room radiography dedicated
Flexible mobile operating performance
Wireless remote exposure, greatly reducing the dose of physician radiation